Leak test of NA58 COMPASS polarized target safety valves using Alcatel leak detector. The valves were connected to the pumping port of the detector with vacuum on the high pressure side of the valve. Leaks through the valve and from ambient air were checked by blowing helium gas on the valves. Detected leak rate in units mbar l/s is given. The vacuum pressure was less than 2 10^-3 mbar. Safety valves for magnet Valve no Pressure Description Leak rate[mbar l/s] PSV25 8844 +1.5 bar was LN2 cooling line not tested PSV23 8845 +1.5 bar was LN2 cooling line not tested PSV11046 11046 +1.75 bar LN2 cooling line not tested PSV11047 11047 +1.75 bar LN2 cooling line not tested PSV65 8842 +0.5 bar Helium dewar not tested PSV66 8841 +1.25 bar Helium dewar not tested Safety valves for dilution cryostat Valve no Pressure Description Leak rate[mbar l/s] VS54 8136 +0.5 bar separator to air <10^-8 VS61 9648 +0.2 bar 4He pumping line to air <10^-8 VS66 2085 +0.5 bar cavity pumping line to air <10^-8 VS42 9647 +0.15 bar 3He pumping line to tanks <10^-8 VS29 2708 +0.5 bar 3He condensation line to tanks <10^-8 outside <10^-5 through VS43 9649 +1.5 bar control valve ref pressure <10^-7 VS71 9650 +1.0 bar 3He service pump exhaust <10^-6 VS99 9651 +2.0 bar LN2 inlet not tested PSV99 8843 +0.5 bar LN2 dewar not tested Thu Mar 12 13:53:21 CET 2009 Jaakko Koivuniemi