Target platform power consumers, "not used" could have been used by SMC magnet but is disconnected now. Wed Sep 7 10:52:18 CEST 2005 Jaakko Koivuniemi Normal network EXD415/HN2 from EXD413/HN2 O2 4xC25 not used O3 4xC63 solenoid power supply 04 4xC40 dipole power supply 05 4xK16A plug-in dilution cryostat evacuation pumps MIV1000+2100 06 2xC16 R 10 plug-ins with C16 and fault current breaker, rack ground floor Jura side 07 2xC16 S 4 plug-ins with C16 and fault current breaker, Saleve side panel ground floor 08 2xC16 T 10 plug-ins SciFi 3 circuit breaker removed 09 2xC40 R not used 010 2xC32 S goes to division box with 2 2xC16 breakers for NMR rack plug and heaters plug Diesel network EAD415/HN2 from EAD211.04/EHN2 O2 2xC16 R 10 plug-ins C16A with fault current breaker, tape and air heaters, Saleve side under platform 03 4xK20A isolation vacuum pumps, Alcatel 2033 rotary and diffusion pump 04 2xC16 S plug-in rotary pump Alcatel 2012 for magnet LN2 line evacuation 05 2xC16 T general purpose plug-in Jura side panel 06 2xD16 T not used Uninterrupted network EOD415/HN2 from EOB210.05/BA82, one phase only O2 2xC16 not used O3 2xC20 not used O4 2xC16 general purpose plug-in Jura side panel 48 V DC network ECJ415/HN2 from ECD2105/HN2 O1 4 A EUJ415/HN2 AUL platform power emergency stop O2 4 A was LN2 transfer junction ECD2105/HN2 from ERB21 210/12 O1 10 A not used O2 10 A not used O3 10 A not used O4 10 A not used O5 10 A ECJ415/HN2 AUL platform power emergency stop O6 10 A PLC T3(020) target platform PLC I/O box O7 10 A not used O8 10 A not used