Minutes from target meeting held November 21 2014 in 892-2-009 These minutes are available also from: http://wwwcompass.cern.ch/compass/detector/target/Minutes/2014/ a) Close evaporator main needle valve on Friday afternoon and start removing gas from mixing chamber with small rotary pump PP3 in the pump room. The evaporator and mixing chamber empty at midnight. Add 0.2 bar of gas and restart to circulate. Stop and bypass helium-4 pumps on Saturday afternoon 16:30. On Monday morning 77 K reached at evaporator and 78 K in mixing chamber. Pump out gas and fill in 0.2 bar of helium. Repeat purge cycle twice. Fill in 0.2 bar of helium and start to circulate. Start to cool down the dilution cryostat. b) On Tuesday morning restart helium-4 pumps. Some liquid on bottom from 16:00. Warm up the LN2 trap on Wednesday morning. Condense helium-4 to mixing chamber on Wednesday 14:00 - 17:30. c) Add TDR1, TDR2, TDR4, TDR5, TDR6, TPO3 and TTH9 thermometer calibration code to 'mulreg.cxx'. An other small program to calculate temperature from helium-4 vapor pressure 'its90he4.cxx'. See https://git.cern.ch/web/ptnmr.git d) Fabrice: test magnet safety system. Small modification done on the voltage measurement cards at different test points of the coil winding. Up to 50 A current was used in the testing. e) The end of the run planning and liquid helium supply were discussed.