Minutes from target meeting held September 12 2014 in 892-2-009 These minutes are available also from: http://wwwcompass.cern.ch/compass/detector/target/Minutes/2014/ a) Reconnect helium-4 pumps cooling water lines. The roots heat exchanger flow normal. The rotary heat exchanger flow can not be tested without the pump running due to automatic water valve inside the pump. Test brut cooling water from the towers too. A lot of pollution found from this water and some of this was removed by opening the water pipes at the flow meter. The pressure regulator can set the pressure up about 4 bar, which is less than the return line pressure and thus there is no water flow. Only the tap water circuit seems to be working normally. b) Norihiro: circulate warm helium gas through the magnet and dilution cryostat from the cold box. Initial oxygen impurity about 100 ppm with some water vapor too. Closing the valves to the filters after helium-4 pumps improved the situation. c) Yuri: measure all 16 trim coil resistances during the cool down. With the typical copper residual resistance ratio the resistance is expected to go down by a factor about 70. d) Kaori: three NMR coils mounted outside of the target cells and two inside. The inside coils are at the ends of the target cells. Soldering the SMA connectors on the coaxial cables. The new NMR rack under the platform on Jura side already in place. e) Norihiro: receive new Siemens PLC modules that can be used to set needle valve positions remotely.