Minutes from target meeting held August 9 2013 in 892-2-009 These minutes are available also from: http://wwwcompass.cern.ch/compass/detector/target/Minutes/2013/ a) The water manifold in the control room was dismounted and reinstalled on the new wall position but now on the Saleve side. The microwave table is waiting for a new support structure to be placed on (the false floor is not good enough). b) Yuri: feed in the microwave cavity with IMPATT diode microwave generator. The microwave frequency counter is used to see the applied frequency and a diode is used to measure the outcoming power from one of the wave guides. When sweeping the frequency from 69.9 GHz to 70.4 GHz clear amplitude maximums at each 3 - 5 MHz (modes) are distinghuised. The density of these modes shows that there is a good distribution of microwave power inside the cavity.